Impedance Measurement
Precision measurement of extremely low impedance values on printed circuit boards requires a configuration that eliminates the effects of test lead impedance and contact resistance. The four-terminal pair coaxial configuration achieves this by extending the shield of the coaxial cables all the way to the device under test. This arrangement ensures that the current flowing through the inner conductors returns entirely through the outer shields, neutralizing the magnetic fields.
Measurement Mechanism
The coaxial current flow prevents mutual inductance between the current source leads and the voltage sensing leads. This suppression of magnetic coupling allows for highly accurate impedance measurements at frequencies extending into the megahertz range. Standard kelvin connections fail at these frequencies because the electromagnetic fields around the leads induce spurious voltages in the sensing loop.
By maintaining the shield return paths, the four-terminal pair connection maintains an stable measurement environment even when long test leads are necessary. This configuration is standard in automated impedance analyzers used to test the capacitance of embedded planar capacitors on multilayer boards.
Calibration Method
System calibration requires open, short, and load standards designed to fit the specific coaxial terminal interface. These standards define the reference plane right at the connection point of the device under test. Residual inductances are mathematically subtracted during the calibration run.
This process ensures that measured impedance changes reflect only the properties of the circuit under test.