Calibration Adjustment
Adjustment of electrical resistance in the power supply leads of a bridge circuit ensures that the voltage reaching the sensors remains at the specified level despite variations in wire length. Technicians apply excitation line trimming to compensate for voltage drops. This procedure prevents the attenuation of the input signal which would otherwise introduce a systematic offset in the output readings of the load cell or pressure transducer.
Resistance Balancing
The process involves adding or removing resistive material from the traces or inserting discrete components to match the impedance requirements of the measurement system. In high precision assembly, excitation line trimming corrects for the ohmic losses inherent in long cable runs or thin copper traces on a circuit board. Physical modification of the circuit path allows the instrument to maintain a stable reference regardless of the distance from the regulator.
Fine tuning of these resistors occurs during the final calibration phase of a multi channel acquisition unit where individual channel variances must be nullified to achieve uniform sensitivity across all inputs.
Error Suppression
Failure to perform this step results in a gain error where the sensitivity of the entire measurement chain is lower than the design specifications. Measurement accuracy relies on the precise delivery of the excitation voltage because any drop in the supply directly scales the resulting output signal. Accurate excitation line trimming provides the necessary headroom for digital signal processing to function within its linear range.