Vector Correction
Microwave measurement techniques use specific transmission line standards to remove systematic errors from a vector network analyzer. High frequency characterization of PCB traces often employs line-reflect-line calibration because it does not require a highly precise match or a perfectly known open circuit. The process relies on the characteristic impedance of the transmission lines themselves to define the reference plane.
Impedance Determination
Measurement accuracy depends on the fabrication of two lines with identical cross sections but different lengths and a single high reflection standard. During a line-reflect-line calibration sequence, the system measures the scattering parameters of the through connection, the reflect standard, the first line and the second line. This sequence allows the software to calculate the propagation constant and the transition errors at the probe tips without assuming the behavior of the connectors.
Because the line impedance defines the reference, the copper width and dielectric height must be tightly controlled during board manufacturing.
Frequency Constraint
Effective use of this method requires that the phase difference between the two lines stays between 20 and 160 degrees at all target frequencies. When the frequency range is too broad for a single pair of lines, the designer must include multiple line lengths on the test coupon to maintain accuracy. The physical requirement limits the available space on the panel for production circuits.
Misalignment of the probes or variations in surface roughness on the pads will introduce phase shifts that invalidate the calibration.