Calibration Procedure
High frequency instrument correction removes systematic errors by measuring four physical standards at the end of the test cables. The solt calibration method uses short, open, load and thru connections to characterize the vector network analyzer. This process establishes a reference plane at the point where the device under test is connected.
By comparing the measured values of these standards against their mathematical models, the instrument calculates error terms for directivity and source match.
Error Correction
Twelve term error models rely on these measurements to account for tracking and crosstalk within the test setup. The open and short standards provide information about the reflection tracking, while the load standard determines the directivity. Connecting the two ports together with a thru standard allows the system to measure transmission tracking and load match.
High precision connectors and clean mating surfaces are essential for achieving repeatable results during this phase.
Standard Requirement
Mechanical standards must be handled with care to avoid damage that would invalidate the calibration. Each standard has a specific coefficient file provided by the manufacturer that describes its parasitic inductance and capacitance. Using the wrong definition for a calibration kit leads to inaccurate impedance measurements and phase shifts.
After the procedure is complete, the analyzer displays the corrected performance of the board or component by mathematically removing the influence of the cables and connectors.